ADFM® Flow Meter
Section 2 Setup and Installation
2-16
Figure 2-14 Connected screen (status window)
Testing the ADFM
Diagnostic tests are selectable from the buttons to the right of
the screen. The results of each test are displayed and recorded in
the station’s Log File.
The Field Test should be performed in the field with the sensor
installed and underwater. The test performs internal ADFM
system checks. Click Field Test to perform the test.
Note
The Bench Test function is for diagnostic purposes and is not
required for routine operation. If for any reason the Bench Test
is
performed, regardless of the time interval between tests, first
retrieve and save all recorded data before it is performed
again. Prior to redeployment, select "Erase Recorder" and
then "Restart."
5. When a connection is established, a status message will
appear in the window; when the status message appears,
select P
ROGRAM
AND
G
O
.
This programs the ADFM with the information and opera-
tional parameters you saved in the Station file, and acti-
vates flow monitoring operation