Operation
3000TB
Teledyne Analytical Instruments
62
6.09 ppm Anlz
1- 136730 30:00
The second line will display three numbers for a few seconds. The
first number (1) is the gain of the amplifier (1 is the highest gain used
for low ppm analysis).
The second number (136730) is the ADC count from the A to D
converter.
The third number (30:00) is the timer in minutes: seconds, and this
is the value that it shows when it is not triggered.
When the timer reaches 00:00, the system alarm is triggered and the
message “
FSEN
“ and “
ERRS
” appear in the upper right corner of the
display instead of “
Anlz
”.
0.01 ppm FSEN
Range: 0 - 100
0.01 ppm ERRS
Range: 0 - 100
To reset the message and the system alarm contacts, the operator
must go back to the Sensor function in the system menu and select
Reset
, then press the
Enter
key. The display will be followed by a
prompt to confirm the reset.
Reset Sensor Fail?
<ENT> Yes <ESC> No
4.10 Valve Box Functions
Three additional functions are available from the System Menu
that are used exclusively with analyzers that interface with a Teledyne
VB Valve Box ot Teledyne Adapter Profibus (TAPA) board. These are:
Serial #
Stream