Operation
3000TB
Teledyne Analytical Instruments
50
Once span settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the
Analyze
mode.
4.4.1.3
C
ELL
F
AILURE
Cell failure in the 3000TB is usually associated with inability to zero
the instrument down to a satisfactorily low ppm reading. When this occurs,
the 3000TB system alarm trips, and the LCD displays a failure message.
#.# ppm Anlz
CELL FAIL/ ZERO HIGH
Before replacing the cell:
Check your span gas to make sure it is within
specifications.
Check for leaks downstream from the cell, where oxygen
may be leaking into the system.
If there are no leaks and the span gas is OK, replace the cell as de-
scribed in chapter 5,
Maintenance
.
4.4.2 Span Cal
The
Span
button on the front panel is used to span calibrate the
analyzer. Span calibration can be performed using the
automatic
mode,
where an internal algorithm compares consecutive readings from the
sensor to determine when the output matches the span gas concentration.
Span calibration can also be performed in
manual
mode, where the
operator determines when the span concentration reading is acceptable
and manually exits the function.
4.4.2.1
A
UTO
M
ODE
S
PANNING
Press
Span
to enter the span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually. Use the ▲/▼ arrow keys to toggle between
AUTO
and
MAN
span settling. Stop when
AUTO
appears, blinking, on
the display.
Span: Settling: AUTO
<ENT> For Next
Summary of Contents for 3000TB
Page 2: ......
Page 3: ......
Page 4: ......
Page 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Page 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......