Trace Oxygen Analyzer
Installation
Teledyne Analytical Instruments
23
3.3.3 Analog Outputs
There are eight DC output signal connectors on the ANALOG
OUTPUTS connector block. There are two connectors per output with
the polarity noted. See Figure 3-5.
Figure 3-5: Analog Output Connections
The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing oxygen,
from 0 V at 0 ppm to 1 V at full scale ppm.
(Full scale = 100% of programmable range.)
0–1 V dc Range ID:
0.25 V = Low Range
0.5 V = Medium Range
0.75 V = High Range
1 V = Air Cal Range
4–20 mA dc % Range: Current increases linearly with increasing
oxygen, from 4 mA at 0 ppm to 20 mA at full
scale ppm. (Full scale = 100% of programmable
range.)
4–20 mA dc Range ID: 8 mA = Low Range
12 mA = Medium Range
16 mA = High Range
20 mA = Air Cal Range