Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
45
Use the ◄►arrow keys to position the blinking over the
Logout
function, and press
Enter
to Log out. The screen will display the
message:
Protected Until
Password Reentered
4.3.5 System Self-Diagnostic Test
The Model 3000TB has a built-in self-diagnostic testing routine.
Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the
test the status of each function is displayed on the screen, either as
OK
or as a number between 1 and 3. (See
System Self Diagnostic Test
in
chapter 5 for number code).
The self diagnostics are run automatically by the analyzer
whenever the instrument is turned on, but the test can also be run by the
operator at will. To initiate a self diagnostic test during operation:
Press the
System
button to start the
System
function.
TRAK/HLD Auto—Cal
PSWD Logout More
Use the ◄►arrow keys to blink
More
, then press
Enter
.
Version Self—Test
Use the ◄►arrow keys again to move the blinking to the
Self–
Test
function. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp — 83