Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
51
Press
Enter
to move to the next screen.
Calib. Holding time
Cal hold: 5 min
This menu allows the operator to set the time the analyzer should
be held in the span mode, after the readings of the analyzer settle. Five
minutes is the default, but it could be adjusted anywhere from 1 to 60
minutes by using the ▲/▼ keys. (A longer cal holding time will result
in a slightly more accurate calibration. Five minutes is appropriate for
most users and applications.)
Press
Enter
to move to the next screen.
Span Val: 000008.00
<ENT>Span <UP>Mod #
Use the ▲/▼ arrow keys to enter the oxygen-concentration mode.
Use the ◄►arrow keys to blink the digit you are going to modify. Use
the ▲/▼ arrow keys again to change the value of the selected digit.
When you have finished typing in the concentration of the span gas you
are using (209000.00 if you are using air), press
Enter
to begin the Span
calibration.
#### ppm
Span
Slope=#### ppm/s
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates
information on
Slope
. Spanning automatically ends when the span
output corresponds, within tolerance, to the value of the span gas
concentration. Then the instrument automatically returns to the analyze
mode.
Summary of Contents for 3000TB
Page 2: ......
Page 3: ......
Page 4: ......
Page 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Page 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......