Operation
3000TB
Teledyne Analytical Instruments
44
A A A A A
<ENT> To Proceed
Use the arrow keys to retype your password and press
Enter
when
finished. Your password will be stored in the microprocessor and the
system will immediately switch to the
Analyze
screen, and you now
have access to all instrument functions.
If all alarms are defeated, the
Analyze
screen appears as:
0.0 ppm
Anlz
Range: 0 — 100
If an alarm is tripped, the second line will change to show which
alarm it is:
0.0 ppm
Anlz
AL—1
Note: If you log off the system using the logout function in the
system menu, you will now be required to re-enter the
password to gain access to Span, Zero, Alarm, and Range
functions.
4.3.4 Logout
The
Logout
function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the
instrument off. By entering
Logout
, you effectively log off the
instrument leaving the system protected against use until the password is
reentered. To log out, press the
System
button to enter the
System
function.
TRAK/HLD Auto—Cal
PSWD Logout More
Summary of Contents for 3000TB