Trace Oxygen Analyzer
Installation
Teledyne Analytical Instruments
19
Installation
Installation of the Model 3000TB Analyzer includes:
1.
Unpacking
2.
Mounting
3.
Gas connections
4.
Electrical connections
5.
Installing the Micro-Fuel Cell
6.
Testing the system.
3.1 Unpacking the Analyzer
Although the analyzer is shipped complete, certain parts, such as
fuses and sensors, are wrapped separately to be installed on site as part
of the installation. Carefully unpack the analyzer and inspect it for
damage. Immediately report any damage or shortages to the shipping
agent.
3.2 Mounting the Analyzer
The Model 3000TB is designed for bulkhead mounting in non-
hazardous environments. There are four mounting lugs—one in each
corner of the enclosure. The outline drawing, at the back of this manual,
gives the mounting hole size and spacing. The drawing also contains the
overall dimensions. Do not forget to allow an extra 1 3/8" for the hinges.
Be sure to allow enough space in front of the enclosure to swing the
door open—a 11 3/4" radius, as shown in Figure 3-2.
All electrical connections are made via cables which enter the
housing through ports in its side. No conduit fittings are supplied. The
installer must provide two 3/4" NPT and two 1" NPT adapters and the
appropriate sealing conduit.
For gas connections, the unit is supplied with 1/8" tube fittings, and
three millimeter adapters are supplied for metric system installations.
Figure 3-1 shows the Mode3l 3000TB in front view with the front
door closed.
Summary of Contents for 3000TB
Page 2: ......
Page 3: ......
Page 4: ......
Page 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Page 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......