![Teledyne 3000TB Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/teledyne/3000tb/3000tb_operating-instructions-manual_1080406033.webp)
Operational Theory
Model 3000TB
Teledyne Analytical Instruments
18
In the presence of oxygen the cell generates a current. A current to
voltage amplifier converts this current to a voltage, which is amplified in
the second stage amplifier.
The second stage amplifier also supplies temperature compensation
for the oxygen sensor output. This amplifier circuit incorporates a ther-
mistor, which is physically located in the cell block. The thermistor is a
temperature dependent resistance that changes the gain of the amplifier
in proportion to the temperature changes in the block. This change is in-
versely proportional to the change in the cell output due to the same
temperature changes. The result is a signal that is temperature
independent. The output from the second stage amplifier is sent to an 18
bit analog to digital converter controlled by the microprocessor.
The digital concentration signal along with input from the control
panel is processed by the microprocessor, and appropriate control
signals are directed to the display, alarms and communications port. The
same digital information is also sent to a 12 bit digital to analog
converter that produces the 4-20 mA DC and the 0-1 VDC analog
concentration signal outputs, and the analog range ID outputs.
Signals from the power supply are also monitored, and through the
microprocessor, the system failure alarm is activated if a malfunction is
detected.
Summary of Contents for 3000TB
Page 2: ......
Page 3: ......
Page 4: ......
Page 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Page 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......