Appendix A: TLA721 Mainframe Power Information
TLA700 Series Logic Analyzer Installation Manual
57
For high power usage combined with low input line voltages (shaded region),
use only the power cord with the 20 A plug. Select the proper fuse based on the
ranges shown in Figure 29.
Po
wer
Consum
pt
ion
(W
at
ts)
800
700
600
500
400
300
200
100
0
Input Line Voltage (Volts AC)
90
110
115
130
150
170
190
210
230
250
20AT Fuse
15AF Fuse
6.3AF Fuse
Power Cord with 20A plug (NEMA 5--20P) ONLY!
Power Cord with 15A plug (NEMA 5--15P)
Power Cord with 20A plug (NEMA 5--20P)
Figure 29: Power cord identification chart
For example, assume that your TLA721 benchtop system consists of four
TLA7Q4 Logic Analyzer Modules and one TLA7E2 DSO Module. Also assume
that you will be operating the mainframe at 90 VAC.
Add 121 Watts (for the DSO Module) + 4 X 75 Watts (for the logic analyzer
modules) + 70 Watts (for the controller) + 100 Watts for the mainframe to come
up with a total value of 591 Watts.
Because the power consumption (at 90 VAC) is approaching the shaded area of
the graph in Figure 29, you should consider using the 20A power cord.
NOTE
. If you are configuring an expansion mainframe, remember to use the
power value for the expansion module instead of the benchtop controller.