Restoring and Reinstalling Software
TLA700 Series Logic Analyzer Installation Manual
53
25.
If the firmware versions do not match, power down the mainframe, remove
the module from the mainframe and update the label.
Use the following procedure to upgrade the firmware on the pattern generator
modules.
1.
Note which modules require new firmware.
If any modules are missing from the System window, you can verify the
status of the modules by doing the following steps:
a.
Select System Properties in the System menu.
b.
Click the Modules tab.
c.
Note the information under the Messages column for any installed
modules. If any modules indicate invalid or similar messages, you must
update the firmware for those modules.
2.
Power down the mainframe and disconnect the power cord.
3.
Remove the modules from the mainframe.
4.
Refer to Figure 28 on page 51 and locate the flash programming pins on the
rear of each module.
5.
Install a jumper on the flash programming pins (use one of the spare jumpers
that came with your pattern generator).
6.
Reinstall the modules in the mainframe.
7.
Reconnect the power cord, and then power on the mainframe.
8.
Wait for the pattern generator application to start.
NOTE
. Any modules with the flash programming jumper installed will not display
in the System window.
9.
Exit the pattern generator application (and the TLA application).
10.
Click Start
→
Programs
→
Tektronix Logic Analyzer
→
TLA Firmware
Loader.
11.
Select the modules that you want to update from the list of modules
displayed in the Supported list box near the top of the window. This selects
the modules for the upgrade.
12.
If you are upgrading more than one module at a time, note the slot location
of the module in the selection box.
Upgrading Firmware on
the TLA7PG2 Pattern
Generator Modules