Measurement Algorithms
Figure 133: Equipment setup for Reset from High Speed test
Packet Parameter Test Equipment Setup
Test Fixture Setup
To set up the test fixture, follow these steps:
1.
Set the S5 switch to the Init position.
2.
Connect the standard USB cable between the Device SQ Init port and the
host port.
3.
Connect the A receptacle (marked DUT) from the Device SQ test port to the
B receptacle device.
4.
Apply the power to the test fixture.
5.
Apply the power to the DUT.
6.
Attach the differential probe near the device connector on the test fixture.