![Tektronix TDSUSB2 Manual Download Page 204](http://html1.mh-extra.com/html/tektronix/tdsusb2/tdsusb2_manual_1078195204.webp)
Measurement Algorithms
The intersection of the four voltage points p1, p2, q1, and q2 gives the Cross-
Over point for Voltage level (y) and time (x). x and y coordinates are obtained by
solving the following two equations:
Where: p1, p2, q1, and q2 are the consecutive data points of a single-ended
signal.
x and y are the Cross-Over coordinates x and y that is given by the intersection of
p1, p2, q1, and q2.
t1 and t2 are the time values for the data points p1 and p2.
190
TDSUSB2 Universal Serial Bus Measurement Package
Summary of Contents for TDSUSB2
Page 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Page 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Page 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Page 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Page 90: ...Tutorial 76 TDSUSB2 Universal Serial Bus Measurement Package...
Page 160: ...Application Examples 146 TDSUSB2 Universal Serial Bus Measurement Package...
Page 230: ...Measurement Algorithms 216 TDSUSB2 Universal Serial Bus Measurement Package...