![Tektronix TDSUSB2 Manual Download Page 189](http://html1.mh-extra.com/html/tektronix/tdsusb2/tdsusb2_manual_1078195189.webp)
Reference
Table 36.Error Code descriptions (cont.)
Miscellaneous
E1000
Invalid device Chirp-K for this
test
As part of the USB2.0 description, a
minimum of 3 Chirp-K and 3 Chirp-J
states are expected after Chirp
Response timing. An absence of this
may lead to incorrect results
Try acquiring the signals again by pressing
Run button
E1001
No SOF found
No SOF in D+ and D-
Power off the DUT, check the probe
connection and try acquiring the signal
E1002
The current signal is not a Reset
From Suspend signal. Acquire
the correct signal and proceed
Acquired signal is not a correct Reset
From Suspend signal. Correct Reset
From Suspend Signal is required to
perform the measurement
Try acquiring the signals again by pressing
Run button
E1003
The current signal is not a
Resume signal acquire the
correct signal and proceed
Acquired signal is not a correct
Resume signal. Correct Resume
signal is required to perform the
measurement
Try acquiring the signals again by pressing
Run button
E1004
The current signal is not a
Suspend signal acquire the
correct signal and proceed
Acquired signal is not a correct
Suspend signal. Correct Suspend
signal is required to perform the
measurement
Try acquiring the signals again by pressing
Run button
E1005
There is no falling edge
transition for this test.
If there is no J state found
Try acquiring the signals again by pressing
Run button
E1006
There is no start of Chirp-K state
for this device
If there is no K state found
Try acquiring the signals again by pressing
Run button
E1007
There is no K State transition
driven by the host controller
If there is no K state found
Try acquiring the signals again by pressing
Run button
E1008
The current signal is not a Reset
From Resume signal. Acquire
the correct signal and proceed
The acquired signal is not a correct
Reset From Resume signal. Correct
Reset From Resume signal is required
to perform the measurement
Try acquiring the signals again by pressing
Run button
TDSUSB2 Universal Serial Bus Measurement Package
175
Summary of Contents for TDSUSB2
Page 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Page 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Page 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Page 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Page 90: ...Tutorial 76 TDSUSB2 Universal Serial Bus Measurement Package...
Page 160: ...Application Examples 146 TDSUSB2 Universal Serial Bus Measurement Package...
Page 230: ...Measurement Algorithms 216 TDSUSB2 Universal Serial Bus Measurement Package...
Page 250: ...Glossary 236 TDSUSB2 Universal Serial Bus Measurement Package...