SPG8000A User Manual
37
c.
Press the up (▲) or down (▼) arrow button to select
SYSTEM : DIAGNOSTICS : POWER SUPPLY 1
. The
second line of the display will list the number of hours the
supply has been the active supply and the backup
supply.
d.
Press the right (►) arrow button to display the number of
temperature
-
weighted hours of the first supply as shown
.
Make a n
ote
of the number of hours of hours
shown.
e.
Press the up (▲) or down (▼) arrow button to select
SYSTEM : DIAGNOSTICS : POWER SUPPLY 2
.
f.
Press the right (►) arrow button to display the number
temperature weighted hours of the second supply. Note the
number of hours of hours shown.
3.
If
the
Power
Supply
module
with
the
greatest
number
of
temperature
weighted hours is not the currently the preferred supply,
then continue
this procedure. Otherwise, you do not need to change the preferred
supply
configuration.
4.
To
change the preferred supply configuration, press
SYSTEM
to
access the SYSTEM
menu.
5.
Press the u
p (▲) or down (▼) arrow button to select
SYSTEM :
POWER SUPPLY PREFERRED SELECT
.
6.
Press the left (◄) or right (►) arrow button to display the supply
you want to be the primary (active) supply: Power Supply 1 or
Power Supply 2. The second line of the display
shows the supply
status:
Active, Preferred
or
Inactive
.
Note
:
When viewed from the front of the instrument, Power Supply 1 is the left
Power Supply module and Power Supply 2 is the right Power Supply module.
This matches the orientation of the PWR1 and PWR2 indicators on the front
panel.
7.
If status of the supply is Inactive, press
ENTER
to change the status
to Active, Preferred. The front
-
panel indicator for the selected supply
should change to bright
green.
Hot-Swap Power Supply Module
This procedure describes how to replace a
p
ower
s
upply module while the
instrument is in operation.
To hot-
swap a
p
ower
s
upply module:
WARNING:
Disconnect the power cord from the Power Supply module
before you remove the module from the instrument. Also, do not connect
a power cord to the Power Supply module while the module is not
Summary of Contents for SPG8000A
Page 53: ...SPG8000A User Manual 39 Figure 7 Removing a Power Supply module...
Page 103: ...SPG8000A User Manual 89 Figure 22 SDI Overlay submenu...
Page 182: ...168 SPG8000A User Manual Figure 61 AES button menu diagram...
Page 190: ...176 SPG8000A User Manual Figure 73 EMBED button menu diagram...
Page 194: ...180 SPG8000A User Manual Figure 77 DOLBY E CONFIG submenu diagram part 2...
Page 197: ...SPG8000A User Manual 183 Figure 81 REF button menu diagram part 2...
Page 200: ...186 SPG8000A User Manual Figure 84 SDI button menu diagram part 2...
Page 211: ...SPG8000A User Manual 197 Figure 96 SDI ANC PAYLOAD submenu diagram part 2...
Page 213: ...SPG8000A User Manual 199 Figure 97 STATUS button menu diagram part 1...
Page 215: ...SPG8000A User Manual 201 Figure 99 SYSTEM button menu diagram part 1...
Page 216: ...202 SPG8000A User Manual Figure 100 SYSTEM button menu diagram part 2...
Page 225: ...SPG8000A User Manual 211 Figure 107 SNMP TRAPS submenu diagram part 2...
Page 240: ...226 SPG8000A User Manual Figure 123 PTP primary master menu diagram part 1...
Page 241: ...SPG8000A User Manual 227 Figure 124 PTP primary master menu diagram part 2...
Page 243: ...SPG8000A User Manual 229 Figure 126 PTP primary slave menu diagram part 2...