P6205 Instruction Manual
3–1
Specifications
This section lists the electrical, environmental, and physical
specifications of the P6205 probe. All specifications are guaranteed
unless labeled “typical”. Typical specifications are provided for your
convenience and are not guaranteed. Specifications marked with the
n
symbol are verified in in the Performance Verification section
beginning on page 4–1.
The electrical characteristics listed in Table 3–1 apply to a probe
calibrated between 20
_
C and 30
_
C. The instrument system must
also be calibrated and operating within the environmental conditions
listed in Table 3–2 on page 3–4.
Following the specification tables you will find a series of graphs
showing typical probe characteristics.
Table 3–1: Electrical characteristics
Characteristic
Description
n
Attenuation
1
10X
±
1.8% at DC into 50
±
0.5% load
Input resistance
2,3
1 M
±
5% at DC
Input capacitance
≤
2 pF
n
Rise time
3
< 467 ps
Bandwidth
3
Derived from rise time (BW = 0.35/t
r
): 750 MHz
Summary of Contents for P6205
Page 4: ......
Page 12: ...Service Safety Summary viii P6205 Instruction Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 24: ...Getting Started 1 8 P6205 Instruction Manual...
Page 25: ...Operating Basics...
Page 26: ......
Page 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Page 33: ...Specifications...
Page 34: ......
Page 41: ...Performance Verification...
Page 42: ......
Page 57: ...Maintenance...
Page 58: ......
Page 65: ...Replaceable Parts...
Page 66: ......
Page 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Page 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...