Performance Verification
4–4
P6205 Instruction Manual
Figure 4–2: Calibrator waveform
5. Store and then recall the waveform.
6. Use the oscilloscope cursors to display the
D
voltage of the
recalled waveform. Measure the low-frequency portion of the
waveform near the back corner. See figure 4–3 below.
Figure 4–3: Cursors placement for
voltage measurement
7. Record the
D
voltage as V
1
.
8. Disconnect the coaxial cable from the oscilloscope calibrator
output and reconnect it to the channel 1 output of the 1103 probe
power supply. Remove the SMA-to-BNC adapter from the
oscilloscope calibrator output.
9. Connect the P6205 probe output to the 1103 power supply
channel 1 input. The P6205 does not use the offset capability of
the 1103 power supply; the offset control setting is not important.
10. Mate the GR connector ends of the 50
W
termination adapter and
the male SMA-to-GR adapter. Connect the mated assembly to the
oscilloscope calibrator output. Insert the probe tip firmly into the
50
W
termination adapter. See Figure 4–4 below.
Summary of Contents for P6205
Page 4: ......
Page 12: ...Service Safety Summary viii P6205 Instruction Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 24: ...Getting Started 1 8 P6205 Instruction Manual...
Page 25: ...Operating Basics...
Page 26: ......
Page 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Page 33: ...Specifications...
Page 34: ......
Page 41: ...Performance Verification...
Page 42: ......
Page 57: ...Maintenance...
Page 58: ......
Page 65: ...Replaceable Parts...
Page 66: ......
Page 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Page 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...