Performance Verification
P6205 Instruction Manual
4–11
Measure the peak negative-going aberration (if any) and the peak
positive-going aberration (if any). The sum of the two is the
peak-to-peak aberration, in percent.
NOTE
. If the waveform front corner appears rounded or rolled-off, set
the left-limit measurement cursor at the first 100% amplitude point
of the waveform to determine aberrations. Refer to Figure 4–9.
Positive aberration
100%
90%
10%
Rise time
20ns
Negative aberration
2
m
s
Peak-to-peak aberration
Figure 4–9: Measuring aberrations
7. Disconnect the coaxial cable from the oscilloscope calibrator
output and reconnect it to the channel 1 output of the 1103 probe
power supply. Remove the SMA-to-BNC adapter from the
oscilloscope calibrator output. See Figure 4–10 below.
8. Connect the P6205 probe output to the 1103 power supply
channel 1 input. The P6205 does not use the offset capability of
the 1103 probe power supply; the offset control setting is not
important.
Summary of Contents for P6205
Page 4: ......
Page 12: ...Service Safety Summary viii P6205 Instruction Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 24: ...Getting Started 1 8 P6205 Instruction Manual...
Page 25: ...Operating Basics...
Page 26: ......
Page 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Page 33: ...Specifications...
Page 34: ......
Page 41: ...Performance Verification...
Page 42: ......
Page 57: ...Maintenance...
Page 58: ......
Page 65: ...Replaceable Parts...
Page 66: ......
Page 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Page 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...