Performance Verification
P6205 Instruction Manual
4–3
Oscilloscope settings (Cont.)
Control
Setting
Trigger source
Internal clock
Trigger mode
Auto
Coupling
DC
Waveform averaging
On
Number of waveforms to average
8
Cursor type
Horizontal bars
Cursor knob resolution
Fine
2. Install one SMA-to-BNC adapter on an SD-24 sampling head
input channel and another adapter on the oscilloscope calibrator
output. Use the 10-inch coaxial cable to connect the oscilloscope
calibrator output to the SD-24 input channel. See Figure 4–1
below.
11801B oscilloscope
Figure 4–1: Attenuation test setup part 1
3. Press SELECT CHANNEL on the sampling head to enable the
corresponding input channel.
4. Adjust the oscilloscope controls as necessary to display a
waveform similar to the one shown in Figure 4–2 below.
Summary of Contents for P6205
Page 4: ......
Page 12: ...Service Safety Summary viii P6205 Instruction Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 24: ...Getting Started 1 8 P6205 Instruction Manual...
Page 25: ...Operating Basics...
Page 26: ......
Page 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Page 33: ...Specifications...
Page 34: ......
Page 41: ...Performance Verification...
Page 42: ......
Page 57: ...Maintenance...
Page 58: ......
Page 65: ...Replaceable Parts...
Page 66: ......
Page 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Page 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...