Performance Verification
4–8
P6205 Instruction Manual
4. Measure and record as system rise time (tr
s
), the rise time of the
displayed pulse.
5. Mate the 50
W
termination adapter to the female BNC-to-GR
adapter. Disconnect the coaxial cable at the oscilloscope and
attach the mated adapters to the coaxial cable. Connect the P6205
probe to the same oscilloscope input and insert the probe tip
firmly into the termination adapter.
6. Save and then recall the waveform.
7. Adjust the oscilloscope controls to display this waveform at the
same vertical scale as the first waveform. Measure and record as
probe-plus-system rise time (tr
p+s
), the rise time of the displayed
pulse.
8. Calculate the probe rise time (tr
p
) using the equation below.
Accuracy of the approximation is limited by oscilloscope system
performance.
t
r
p
+
t
2
r
p
)
s
– t
2
r
s
Ǹ
The calculated probe rise time should be less than 467 ps. If the rise
time value exceeds this limit, use the procedure beginning on
page 4–5 to more accurately determine the probe rise time.
Bandwidth Approximation
You can calculate the approximate bandwidth from the rise time
measurement with the following equation:
BW
+
.35
t
r
NOTE
. This calculation does not constitute a valid performance
verification criteria.
Summary of Contents for P6205
Page 4: ......
Page 12: ...Service Safety Summary viii P6205 Instruction Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 24: ...Getting Started 1 8 P6205 Instruction Manual...
Page 25: ...Operating Basics...
Page 26: ......
Page 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Page 33: ...Specifications...
Page 34: ......
Page 41: ...Performance Verification...
Page 42: ......
Page 57: ...Maintenance...
Page 58: ......
Page 65: ...Replaceable Parts...
Page 66: ......
Page 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Page 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...