Performance Verification
4–10
P6205 Instruction Manual
3. Press SELECT CHANNEL on the sampling head to enable the
corresponding input channel.
4. Adjust the oscilloscope controls as necessary to display a
waveform similar to the one shown in Figure 4–8 below.
Figure 4–8: Aberrations waveform
5. Use the oscilloscope rise time filter capability to limit the
waveform rise time to 100 ps. Use the waveform Vertical
Description menu to set the vertical description as follows:
Filter(M1,100E-12)
NOTE
. The filter description above applies to an SD-24 sampling
head installed in the 11801B mainframe M1 plug-in compartment; if
your sampling head is installed in a different compartment, select the
corresponding mainframe designator.
6. Use the oscilloscope measurement capability to determine the
peak-to-peak amplitude of the filtered pulse. Measure, calculate,
and record (in percent) the system aberrations (abs
s
) in the first
4 ns of the waveform.
NOTE
. Part one of the aberrations specification applies to the first
20 ns of the waveform. Significant aberrations are usually complete-
ly damped beyond approximately 4 ns.
Summary of Contents for P6205
Page 4: ......
Page 12: ...Service Safety Summary viii P6205 Instruction Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 24: ...Getting Started 1 8 P6205 Instruction Manual...
Page 25: ...Operating Basics...
Page 26: ......
Page 32: ...Operating Basics 2 6 P6205 Instruction Manual...
Page 33: ...Specifications...
Page 34: ......
Page 41: ...Performance Verification...
Page 42: ......
Page 57: ...Maintenance...
Page 58: ......
Page 65: ...Replaceable Parts...
Page 66: ......
Page 72: ...Replaceable Parts 6 6 P6205 Instruction Manual Figure 6 2 P6205 probe optional accessories...
Page 74: ...Replaceable Parts 6 8 P6205 Instruction Manual...