Calibration—Type P6042
c.
Connect
the P6042
probe to the 10 Ω loop.
Lock the
slider.
d. Connect the
1 X
probe across the 10 Ω resistor. See Fig.
6-9.
e. Switch Triggering
Mode to Trig, Source to Norm, Time/
cm to 50 ms,
and display
to Single Sweep. Set High Freq
3
dB
Point to 500 kHz. While operating the Reset lever, turn
Triggering
Stability counterclockwise
until the Ready light
stays
on. Push Upper and Lower Store buttons.
f.
Reach
inside the front-panel probe
holder and close the
probe-operated
switch (simulating the probe inserted into the
holder). While
holding
the switch closed, push the DEGAUSS
lever (hold
for the
duration of the sweep) and release.
g.
Check the
stored display for a four-centimeter (0.4 V)
damped
waveform
lasting approximately 200 milliseconds.
See Fig.
6-10.
h.
Remove
the
probes from the 10 Ω loop, place the P6042
probe
in the holder (lock
the slider), and remove the 1 X
probe
from the oscilloscope input.
Push Upper
and Lower
Screen
Storage Off
buttons and switch display to Norm.
9.
Adjust
CURRENT/DIV BALANCE
Range
(R16
or R17)
a.
Reconnect
the P6042 OUTPUT, terminated in 50 Ω, to
the
oscilloscope
input. See Fig. 6-6.
b.
Set
Volts/cm
to 50 mV, Time/cm to 50
µs,
and
Trig
gering
Mode to
Auto. Switch
input coupling to Gnd,
posi
tion
the trace on the center graticule line, and return to DC.
c.
With the CURRENT/DIV switch
set to
1 A, degauss the
probe
and center
the trace with
the OUTPUT DC LEVEL con
trol.
d. Switch CURRENT/DIV to 1 mA and
again degauss the
probe.
e. Adjust
either R16 or
R17 to position the trace near the
center
graticule line. (Leave
the remaining adjustment count
erclockwise.)
f.
Degauss
the
probe several times and check to see that
the
trace
returns to within one centimeter of the prevous
position.
g. Switch
CURRENT/DIV to 10 mA and check range of
CURRENT/DIV BALANCE for
>5 cm. Switch to 1 mA, degauss
probe,
and return trace to center graticule
line with CUR
RENT/DIV
BALANCE.
10.
Adjust Gain
(R107)
a.
Set Time/cm to
1 ms,
and Volts/cm to .2 volt.
b.
Preset
the square-wave generator control as follows:
Symmetry
Midrange
Mode
Hi
Amplitude
c.
Remove the cable from
the P6042
OUTPUT and con
nect
to the
square-wave generator Hi Amplitude output,
using
the GR-to-BNC
adapter.
See Fig. 6-11.
d. Adjust the
square-wave generator Amplitude for 5
centimeters of deflection
(1
volt).
e. Remove the GR-to-BNC
adapter and
connect the cable
to
the P6042 OUTPUT. Connect the GR 50 Ω terminating
current
loop
to the square-wave generator output (producing
a 20
mA square wave through
the current loop).
f.
Set
CURRENT/DIV to 5 mA and Volts/cm
to 50 mV.
Degauss
the probe and connect to the current loop. See Fig.
6-12.
g.
Adjust
R107 for 4 centimeters of deflection.
11.
Check
CURRENT/DIV Attenuation
a.
With
the test
equipment set as in the previous step,
change
CURRENT/DIV to 10 mA and check for 2 centimeters
of
deflection, ±3%.
b.
Remove the
current loop from the square-wave genera
tor
and
disconnect the cable from the P6042 OUTPUT.
c. Connect
the
test equipment as in Fig. 6-11.
d.
Switch
Volts/cm
to 2 volts and adjust square-wave
generator Amplitude for
5 centimeters of deflection (10 volts).
e.
Connect
the test equipment is in Fig. 6-12.
f.
Change Volts/cm to
50 mV and CURRENT/DIV to 50
mA.
Check for 4 centimeters of deflection,
±3%.
g.
Switch
CURRENT/DIV to .1 A and check for 2 centi
meters
of deflection, ±3.
h.
Switch
CURRENT/DIV to .2 A, ,5A and 1A positions
and
check for deflection of 1, 0.4 and 0.2 centimeter,
respectively.
(The 1
A position may
be checked for 0.5
centimeter
with
the Volts/cm switch set to 20 mV.) This step
is
not intended as an accurate check of the attenuator toler
ance, but
as
a check for obvious open-circuits conditions in
the
switch.
6-9
Summary of Contents for P6042 PROBE
Page 3: ......
Page 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Page 7: ...Characteristics Type P6042 1 3...
Page 8: ...Characteristics Type P6042 1 4...
Page 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Page 28: ...NOTES...
Page 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Page 46: ...NOTES...
Page 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Page 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Page 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Page 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Page 68: ...FIG 3 AMPLIFIER PO...
Page 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Page 70: ...FIG 4 CABINET...
Page 71: ...FIG 4 CABINET P6042 PROBE...