Operating
Instructions
—Type P6042
through
the holes in the cabinet.
The
clearance provided by
the feet on
the bottom of the instrument should be
main
tained.
Provide at least three inches of clearance at the sides
and
top
(more if possible). The instrument can be operated
where
the ambient air temperature is between 0°C and
+50°C.
Controls
and Connectors
Fig.
2-2 shows the front-panel controls and connectors on
the
P6042 and describes the function of each.
Installation
When shipped
from the factory, the P6042 is ready to be
used
with either
a 50-ohm system or a high-impedance
oscilloscope
and
has been factory calibrated to meet the
specifications listed
in Section 1. Due to normal differences
in
oscilloscope input characteristics, however, the high-fre
quency
response of the probe should
be checked on the
system
with which it
is to be used before making critical
high-frequency
current measurements.
If the
instrument is to be operated with a
high-impedance
oscilloscope connect
the 50-ohm termination
(011-0049-00),
supplied in the accessory
kit, to
the input connector of the
oscilloscope.
1.
Connect the
P6042
OUTPUT to the input of the oscillo
scope (or termination), using
a 50-ohm
coaxial cable.
2.
Turn the oscilloscope and P6042 power on and allow five
minutes
warm-up time (at 25°C).
3. Set the oscilloscope controls
as follows:
Volts/div
50
mV
Variable
(Volts/div)
Calibrated
Input
Coupling
Ground
4.
Set the P6042
controls
as follows:
OUTPUT
DC LEVEL
Midrange
CURRENT/DIV
BALANCE
Midrange
CURRENT/DIV
1
A
5.
Center
the trace vertically on the CRT, then switch input
coupling
to DC.
6.
Push
the thumb-controlled portion of the probe into the
locked
position.
When the slider is in the
locked position the
shield
around the transformer core is grounded, the PROBE
UNLOCKED
light is
extinguished, and the amplifier is ready
for
operation.
7.
Place
the probe
in the
front-panel receptacle. Momen
tarily
depress
the
DEGAUSS
lever
and
release. (Time required
for
probe degaussing is
200 milliseconds.)
NOTE
To
remove
any magnetic flux present in the probe
transformer
core, always degauss the probe after
Fig.
2-3. P6042
Currant
Probe
connected
to
Type
106
Square-Wave
Generator
output.
2-3
Summary of Contents for P6042 PROBE
Page 3: ......
Page 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Page 7: ...Characteristics Type P6042 1 3...
Page 8: ...Characteristics Type P6042 1 4...
Page 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Page 28: ...NOTES...
Page 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Page 46: ...NOTES...
Page 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Page 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Page 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Page 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Page 68: ...FIG 3 AMPLIFIER PO...
Page 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Page 70: ...FIG 4 CABINET...
Page 71: ...FIG 4 CABINET P6042 PROBE...