Calibration
—Type P6042
Fig.
6-18.
Test
setup
for
step
15.
15.
Check
Aberrations
a.
Remove
the current loop and probe from the
square
wave generator and
connect to pulse generator output. Con
nect
the
delay cable to the pulse generator charge line con
nectors,
using the two lengths of RG-8A/U. See Fig. 6-18.
b.
Set
pulse generator Voltage
Range to 5.0, Pulse
Polarity
to
—
, and turn power
on. Set test oscilloscope Volts/cm to
.2
volt, Time/cm to .2
µs and turn
X10
Mag on.
c.
Adjust pulse
generator Amplitude and test oscillocope
Trig
Level Io obtain a 4-centimeter display. (Use the viewing
hood
if
necessary).
d. Check
aberrations for
<3%, peak
to peak.
e.
Remove the
probe from the pulse generator and place
in the holder.
Turn X10 Mag off.
NOTES
6-15
Summary of Contents for P6042 PROBE
Page 3: ......
Page 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Page 7: ...Characteristics Type P6042 1 3...
Page 8: ...Characteristics Type P6042 1 4...
Page 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Page 28: ...NOTES...
Page 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Page 46: ...NOTES...
Page 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Page 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Page 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Page 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Page 68: ...FIG 3 AMPLIFIER PO...
Page 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Page 70: ...FIG 4 CABINET...
Page 71: ...FIG 4 CABINET P6042 PROBE...