Performance
Check—Type
P6042
Constant
Amplitude Signal Generator
Frequency
Range
Amplitude
Amplitude
Range
Power
50
kHz
Only
5
.5 -
5V
On
c.
Degauss
the
probe
and connect to the current loop.
d. Switch
test oscilloscope input coupling to DC and adjust
CURRENT/DIV
BALANCE
to center the display on the CRT.
Trigger
the sweep.
e. Increase the
signal generator Amplitude until
wave
form
distortion occurs.
Check for an undistorted dynamic
range of
at
least + and —1 cm
(> + and —500 mV at
probe
amplifier output). See Fig. 5-5.
Fig.
5-5.
Dynamic
range
of
P6042,
test
oscilloscope
set
to
0.5
V/div.
11.
Check
Bandwidth
a.
Set Volts/cm to .2
and adjust signal generator Ampli
tude
for
5 centimeters of deflection.
b. Increase the
frequency of the constant amplitude signal
generator
until the amplitude of
the display
reduces from 5
centimeters to 3.5 centimeters. See Fig. 5-6.
Fig.
5-6. Bandwidth
of
P6042 (A)
50
kHz
display
<B)
50
MHz
display.
c. Check
the frequency setting of the signal
generator for
>50
MHz.
d. Calculate the risetime
of
the
probe form the measured
bandwidth.
5-5
Summary of Contents for P6042 PROBE
Page 3: ......
Page 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Page 7: ...Characteristics Type P6042 1 3...
Page 8: ...Characteristics Type P6042 1 4...
Page 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Page 28: ...NOTES...
Page 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Page 46: ...NOTES...
Page 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Page 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Page 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Page 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Page 68: ...FIG 3 AMPLIFIER PO...
Page 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Page 70: ...FIG 4 CABINET...
Page 71: ...FIG 4 CABINET P6042 PROBE...