![Tektronix P6042 PROBE Instruction Manual Download Page 39](http://html.mh-extra.com/html/tektronix/p6042-probe/p6042-probe_instruction-manual_1078086039.webp)
Calibration—Type P6042
Fig.
6-13.
Test
setup
for
step
12.
i.
Remove the
probe from the current
loop and connect
to
the test oscilloscope 5 mA current loop. Set the Calibrator
switch
to the 5 mA square-wave position.
j. Set
Volts/cm
to 50 mV, CURRENT/DIV to 1
mA, and
degauss
the
probe.
Check for 5 centimeters of deflection.
k.
Change CURRENT/DIV to 2
mA and check for 2.5 centi
meters
of deflection.
NOTES
6-11