Type
P6042
SECTION
5
PERFORMANCE
CHECK
Change
information,
if any, affecting this section is found at the rear of the manual.
General
Information
This
section of the manual provides a means of checking
the
performance
of the
Type P6042 Current Probe.
It is
intended
to
check the calibration of
the instrument without
the
need
for performing the complete calibration procedure.
Failure
to meet the requirements given in this procedure
indicates
the
need for
internal checks or adjustments. The
user should
refer to the Calibration section of
this manual.
Performance
of the P6042
should be checked after each
1000
hours
of operation and at least
once every six months
to
assure that the instrument is operating correctly
and
accurately.
The step
by
step
instructions of
the procedure furnish an
orderly
approach to the isolation of possible malfunctions,
and
thus serve
as an aid in troubleshooting and repairing
the
probe. Any maintenance
that is known to be needed
should
be performed
before starting
the procedure. If any
trouble becomes
apparent during the performance check,
this should also
be
corrected and the procedure started
over. Repair and
servicing information is given
in the Main
tenance
section of this manual.
Equipment Required
The following
(or equivalent) items of equipment
are re
quired
to check the performance of the P6042 Current Probe.
If substitute
equipment
is used, it must equal or exceed the
given
requirements
in order to check the P6042 to the given
accuracy.
1. Storage
oscilloscope, Tektronix Type 549 with Tektronix
Type
1A7
vertical plug-in unit.
Minimum alernate require
ments:
Sweep
rates
from 50 ms/cm to 50 µs/cm; single
sweep
display; vertical
deflection factors from 500 µV/cm to
0.5V/cm;
vertical bandwidth
selectable: DC
to 500 kHz, and
0.1
Hz to 100 Hz; 5 mA calibrated current loop.
2. High-frequency
test
oscilloscope, Tektronix Type 647A
with
Type 10A2A and
11B2A
plug-in units. Minimum alter
nate
requirement: Vertical
deflection factors from
2V/cm to
20 mV/cm; sweep
rates
from 1 ms/cm to .02 µs/cm(with
X10
magnifier).
Vertical bandwidth
of DC to 100 MHz. A
single
vertical
input channel is
used.
3.
Square-wave generator, Tektronix Type
106. Minimum
requirements: Output amplitude of 10 V into 50 Ω, 200 mA;
output
repetition rate of
1 kHz.
4.
Constant
amplitude signal generator, Tektronix Type
191.
Minimum requirements: Output amplitude of 5 V into
50 Ω,
100 mA; variable frequency to 50 MHz.
5. Pulse
generator,
Tektronix Type 109. Minimum require
ments:
Risetime of 0.25 ns;
adjustable amplitude; 50 Ω output.
6.
Delay Cable,
Tektronix Type 113. Requirements: 60 ns
delay;
characteristic impedance of 50 Ω.
7. 1x probe
with
BNC
connector, P6011, Tektronix Part
No.
010-0193-00.
8. GR
50 Ω
terminating
current loop, Tektronix Part No.
067-0559-00.
9.
Two GR
50 Ω 10X
attenuators,
Tektronix Part No.
017-0078-00.
10. GR-to-BNC
adapter, Tektronix
Part No. 017-0063-00.
11. Two
5 ns lengths of RG-8A/U with GR connectors, Tek
tronix
Part No. 017-0502-00.
12.
Rectangular viewer,
for use with the Type 647A
test
oscilloscope, Tektronix
Part
No. 016-0039-00.
13.
Resistor, 10 Ω,
±10%, 1/4 W, Tektronix Part No.
316-
0100-00.
NOTE
The
high-frequency performance
checks for the
P6042
Current Probe should be
made with a DC
to
100 MHz test
oscilloscope.
Using a lower band
width oscilloscope may
result in an overall system
bandwidth lower
than
50 MHz. Using a higher
bandwidth
oscilloscope may cause aberrations in
excess
of the specified
3%.
PRELIMINARY
PROCEDURE
1.
Set
the
Voltage and Range Selector switches on the
rear
of the
P6042 to the appropriate
positions to match the
line
voltage
being
used. (Refer to the Operating Instructions
section
of this
manual.)
2. Connect
the P6042 to the power source and turn the
POWER
switch ON. Allow five minutes warmup time at
25°
C.
3.
Place
the
selectable bandwidth plug-in unit in the stor
age
oscilloscope.
Connect the
IX probe to the +Input
connector.
Connect a ground lead to the probe.
4.
Preset
the controls as follows:
Storage
Oscilloscope
Enhance
Mode
Off
Upper
Screen Storage
Off
Lower
Screen Storage
Off
Screen Selector
Full
Auto
Erase
Off
Horizontal
Display
A
Mag
Off
Norm-Single Sweep
Norm
A
Time/cm
50
mS
5-1
Summary of Contents for P6042 PROBE
Page 3: ......
Page 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Page 7: ...Characteristics Type P6042 1 3...
Page 8: ...Characteristics Type P6042 1 4...
Page 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Page 28: ...NOTES...
Page 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Page 46: ...NOTES...
Page 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Page 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Page 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Page 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Page 68: ...FIG 3 AMPLIFIER PO...
Page 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Page 70: ...FIG 4 CABINET...
Page 71: ...FIG 4 CABINET P6042 PROBE...