Maintenance—Type
P6042
Probe
Cable
Replacement
If
the probe cable becomes defective, it may be replaced
as
follows:
1.
Remove
the probe from the cable as described pre
viously under “Probe Disassembly”. (Unsolder the two con
nections
to the bottom half
of the probe assembly.)
2. Remove
the
amplifier cabinet
and unsolder the cable
connections
to
the DEGAUSS
switch and the decoupling
board.
3.
Loosen the cable clamp on the back
of the front sub
panel
and
straighten the cable.
4. Pull the cable
through
the front-panel grommet and
remove
the probe strain relief boot from the cable.
To
replace the
cable, reverse
the above procedure. The
probe
strain
relief boot must be placed on the
new cable
before the
cable is reconnected to the amplifier, due to the
size of
the transformer
receptacle on the probe end of the
cable. Refer
to Fig. 4-3 for
color code of decoupling board
connections.
4-6
Summary of Contents for P6042 PROBE
Page 3: ......
Page 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Page 7: ...Characteristics Type P6042 1 3...
Page 8: ...Characteristics Type P6042 1 4...
Page 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Page 28: ...NOTES...
Page 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Page 46: ...NOTES...
Page 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Page 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Page 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Page 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Page 68: ...FIG 3 AMPLIFIER PO...
Page 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Page 70: ...FIG 4 CABINET...
Page 71: ...FIG 4 CABINET P6042 PROBE...