P5150 Adjustments
P5150 Adjustments
These procedures describe how to make low- and high-frequency compensation adjustments to the P5150 probe. Use
the equipment listed below for the test setup.
Table 3: Recommended test equipment
Equipment
Minimum requirements
Recommended test equipment
1
Oscilloscope
2
≥
200 MHz bandwidth (
fl
oating input)
≥
500 MHz bandwidth
(ground-referenced input)
Tektronix TPS2024B or THS3024
Tektronix TDS3054C
Pulse generator
≤
500 ps rise time,
≥
100 ns pulse width,
≥
3 V
p–p
into 50
Ω
Picosecond Labs 2600C or
Fluke 9500B with 9530 active head
Termination, 50
Ω
feedthrough
50
Ω
male-to-female BNC
011-0049-xx
Adapter, BNC
Female-to-female barrel
103-0028-xx
Adapter, probe tip-to-BNC
Probe tip-to-male BNC
013-0291-xx
BNC cable, precision
50
Ω
, 36 in
012-0482-xx
Adjustment tool
Insulated, straight edge
003-1433-xx
Screwdriver
Flat blade
1
Nine digit part numbers (xxx-xxxx-xx) are Tektronix part numbers
2
P5150 probes are intended for use with oscilloscopes that have isolated (
fl
oating) inputs. The Tektronix TPS2024B and THS3024 oscilloscopes
are the highest bandwidth
fl
oating-input oscilloscopes currently available. If you do not have access to one of these oscilloscopes, use the
recommended ground-referenced model.
Access the Adjustments
To gain access to the adjustments, you must slide the cover off of the probe body. Insert a thin
fl
at-blade screwdriver into the
slot on the top of the compensation box and slide back the cover. (See Figure 13.)
Figure 13: Depress the tab and slide back the cover
P5100A & P5150 High Voltage Probes Technical Reference