P5100A Adjustments
Gain Adjustment (DC Accuracy)
1.
Connect the probe as shown in the
System DC Attenuation Accuracy Check
setup. (See Figure 2 on page 3.)
2.
Set the oscilloscope to:
vertical to 1 V/division
horizontal to 20 ms/division
averaging to 64
3.
Set the calibration generator to:
output frequency = 10 Hz
5 V
p-p
square wave output
output into 1 M
Ω
load
4.
Locate R9 (the DC gain adjustment).
WARNING.
To reduce risk of electric shock, use only the provided insulated adjustment tool when making adjustments.
Hazardous voltages may be present inside the compensation box.
Figure 7: R9 (DC gain) adjustment location
5.
Adjust the DC gain (R9) to a probe output level of 5 V, ±0.088 V, as displayed on the oscilloscope.
6.
Record the output level in the outgoing column of the
System DC Attenuation Accuracy
check in the test record.
8
P5100A & P5150 High Voltage Probes Technical Reference