P5100A Adjustments
Figure 10: Areas of the waveform affected by adjustments R1, R2, and R3
First adjust R1 to
fl
atten out to 20-30 ns.
Next, adjust R3 for the 10-20 ns area.
Adjust R2 for the last ~5 ns. If the overshoot is too great, then adjust R2 to roll off the front corner and readjust R3
followed by R1. The procedure is done when the 3 adjustments have been optimized for the best overall response.
5.
Reassemble the probe. (See page 12,
Reassemble the Probe
.)
P5100A & P5150 High Voltage Probes Technical Reference
11