P5100A Adjustments
High Frequency (HF) Adjustment
1.
Connect the test setup as shown.
Figure 9: Setup to adjust HF response
2.
Set the output level of the pulse generator to 5 volts (20 dB) and the pulse width to 100 ns.
3.
Set the oscilloscope to:
vertical to 1 V/division
horizontal to 4 ns/division
averaging to 64
input impedance to 1 M
Ω
4.
Adjust the HF response as follows:
10
P5100A & P5150 High Voltage Probes Technical Reference