P5100A Performance Veri
fi
cation
P5100A Performance Veri
fi
cation
Use these performance veri
fi
cation procedures to verify that the probe meets the warranted speci
fi
cations. These
procedures should be performed in the order given using a DPO4104B oscilloscope with the equipment recommended in
the following table. (See Table 1.)
WARNING.
The following instructions are for use by quali
fi
ed service personnel only. To avoid electrical shock, do not
disassemble or maintain the probe while it is connected to a signal source other than those speci
fi
ed in these procedures.
NOTE.
The probe must be low-frequency compensated before performance can be veri
fi
ed.
If your probe fails to qualify under these conditions, service-level adjustments (DC gain and high-frequency compensation)
may be required. These adjustments must be done by a trained service person.
Test Equipment
Table 1: Recommended test equipment
Equipment
Minimum requirements
Recommended test equipment
1
Oscilloscope
≥
500 MHz BW, 1 M
Ω
±1% input R
Tektronix DPO4104B
Calibration generator
≤
0.5% accuracy, 10 Hz square wave,
5 V
p–p
output
Fluke 9500B with 9530 active head
Adapter, BNC
Female-to-female barrel
103-0028-xx
Adapter, probe tip-to-BNC
Probe tip-to-male BNC
013-0291-xx
Pulse generator
2
≤
500 ps rise time,
≥
100 ns pulse width,
≥
3 V
p–p
into 50
Ω
Picosecond Labs 2600C or
Fluke 9500B with 9530 active head
Termination, 50
Ω
feedthrough
2
50
Ω
male-to-female BNC
011-0049-xx
BNC cable, precision
2
50
Ω
, 36 in
012-0482-xx
Adjustment tool
2
Insulated, straight edge
003-1433-xx
Screwdriver
2
Flat blade
Knife
2
Razor edge
1
Nine digit part numbers (xxx-xxxx-xx) are Tektronix part numbers
2
Required for adjustment procedures only
Test Setup
1.
Turn on the oscilloscope.
2.
Connect the probe to any channel (1–4) of the oscilloscope.
3.
Turn on the generators and let the probe and test equipment warm up for 20 minutes.
4.
Make a copy of the test record to tabulate the test results. (See Table 2 on page 4.)