Section 5: Command reference
Model 2461 Calibration and Adjustment Manual
5-2
2461-905-01 Rev. A / February 2016
:CALibration:ADJust:DATE
This command stores the user-specified adjustment date.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:ADJust:DATE <year>, <month>, <day>
:CALibration:ADJust:DATE?
<year>
The year the instrument was adjusted;
1995
to
2094
<month>
The month the instrument was adjusted;
1
to
12
<day>
The day the instrument was adjusted;
1
to
31
Details
Use this command to store the adjustment date in the instrument nonvolatile memory. Calibration
constants must be unlocked with the
:CALibration:UNLock
command before sending this
command.
You must send the
:CALibration:SAVE
command after making changes, or your changes will not
be saved through a power cycle.
The
year
,
month
, and
day
parameters must be delimited by commas.
You can use the query form of this command to get the last saved adjustment date.
Example
:CAL:ADJ:DATE 2015, 11, 18
:CAL:SAVE
:CAL:ADJ:DATE?
Set the adjustment date to November 18, 2015.
Save the adjustment date.
Query the adjustment date.
Output:
2015,11,18
Also see
(on page 5-9)
(on page 5-10)