Section 5: Command reference
Model 2461 Calibration and Adjustment Manual
5-12
2461-905-01 Rev. A / February 2016
:CALibration:VERify:DATE
This command stores the user-specified calibration verification date.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:VERify:DATE <year>, <month>, <day>
:CALibration:VERify:DATE?
<year>
The year the instrument was verified;
1995
to
2094
<month>
The month the instrument was verified;
1
to
12
<day>
The day the instrument was verified;
1
to
31
Details
Use this command to store the calibration verification date in the instrument nonvolatile memory.
Calibration constants must be unlocked with the
:CALibration:UNLock
command before sending
this command.
You must send the
:CALibration:SAVE
command after making changes, or your changes will not
be saved through a power cycle.
The
year
,
month
, and
day
parameters must be delimited by commas.
You can use the query form of this command to get the last saved calibration verification date.
Example
:CAL:VER:DATE 2015, 11, 18
:CAL:SAVE
:CAL:VER:DATE?
Set the calibration verification date to November
18, 2015.
Save the calibration verification date.
Query the calibration verification date.
Output:
2015,11,18
Also see
(on page 5-9)
(on page 5-10)