Section 5: Command reference
Model 2461 Calibration and Adjustment Manual
5-4
2461-905-01 Rev. A / February 2016
Example
CAL:ADJ:SENS +2.00003
Adjust the 2 V sense range.
Also see
:CALibration:ADJust:SENSe:DATA?
(on page 5-4)
(on page 5-10)
:CALibration:ADJust:SENSe:DATA?
This command queries the calibration constants for the active sense range.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
CALibration:ADJust:SENSe:DATA?
Details
This command requests the calibration constants for the active range of the sense function.
Constants returned are the positive full-scale value, the positive zero value, the negative full-scale
value, and negative zero value.
The returned constants are in comma-delimited ASCII floating-point format.
Example
:SENS:FUNC "CURR:DC"
:SENS:VOLT:DC:RANG 0.2
:CAL:ADJ:SENS:DATA?
Select the DC current sense function.
Select the 200 mV range.
Request the calibration constants.
Example output:
-7.058334E-06,-1.790742E-10,-
7.058321E-06,+0.00
Also see
(on page 5-3)