Model 2461 Calibration and Adjustment Manual
Section 3: Calibration
2461-905-01 Rev. A / February 2016
3-11
It may not be possible to set the resistance calibrator to the specified value. Use the closest possible
setting and modify reading limits accordingly.
5. Verify that the Model 2461 resistance reading is within the limits in the following table.
Ohms measurement accuracy limits
Model 2461 range
Calibrator resistance*
2 Ω
1.9 Ω
20 Ω
19 Ω
200 Ω
190 Ω
2 kΩ
1.9 kΩ
20 kΩ
19 kΩ
200 kΩ
190 kΩ
2 MΩ
1.9 MΩ
20 MΩ
19 MΩ
200 MΩ
100 MΩ
* Nominal resistance value. Calculate specification tolerance and upper and
lower test limits using actual resistance values.