Model 2461 Calibration and Adjustment Manual
Section 4: Adjustment
2461-905-01 Rev. A / February 2016
4-13
Current range calibration adjustment commands
Step
Command/procedure*
Description
1
:SOUR:CURR:RANG <Range>
Select the source range.
2
:SOUR:CURR -<Range>
Establish negative polarity.
3
DMM reading 1.**
Read the actual output value on the DMM.
4
CAL:ADJ:SOUR <DMM_Reading>
Adjust the source function negative full scale.
5
:SYSTem:ERRor?
Check Model 2461 for errors.
6
CAL:ADJ:SENS <DMM_Reading>
Adjust the sense function negative full scale.
7
:SYSTem:ERRor?
Check Model 2461 for errors.
8
:SOUR:CURR 0.0
Set the output to 0 A.
9
DMM reading 2.**
Read the actual output value on the DMM.
10
:CAL:ADJ:SOUR <DMM_Reading>
Adjust the source function negative zero.
11
:SYSTem:ERRor?
Check Model 2461 for errors.
12
:CAL:ADJ:SENS <DMM_Reading>
Adjust the sense function negative zero.
13
:SYSTem:ERRor?
Check Model 2461 for errors.
14
:SOUR:CURR +<Range>
Establish positive polarity.
15
DMM reading 3.**
Read the actual output value on the DMM.
16
:CAL:ADJ:SOUR <DMM_Reading>
Adjust the sense function positive full scale.
17
:SYSTem:ERRor?
Check Model 2461 for errors.
18
:CAL:ADJ:SENS <DMM_Reading>
Adjust the source function positive full scale.
19
:SYSTem:ERRor?
Check Model 2461 for errors.
20
:SOUR:CURR 0.0
Set the output to 0 A.
21
DMM reading 4.**
Read the actual output value on the DMM.
22
:CAL:ADJ:SOUR <DMM_Reading>
Adjust the source positive zero.
23
:SYSTem:ERRor?
Check Model 2461 for errors.
* Perform the complete procedure for each range, where <Range> = 1e-6, 10e-6, 100e-6, 1e-3, 10e-3, 100e-3, and 1.
** The
<DMM_Reading
n
>
parameter is the DMM reading from the previous step.