Section 4: Adjustment
Model 2461 Calibration and Adjustment Manual
4-12
2461-905-01 Rev. A / February 2016
5. Adjust each current range using the procedure summarized in the following table. Do the
following for each range:
Send the
:SOURce:CURRent:RANGe
command to select the source and sense range to
adjust. For example, for the 1 mA range, the command is:
:SOUR:CURR:RANG 1e-3
Program the source to output the negative full-range value using the
:SOURce:CURRent
command. For example:
:SOUR:CURR -1e-3
Record the DMM reading in the "Current calibration adjustment data" table in
(on page 4-4).
Use the DMM reading as the parameter for the
:CALibration:ADJ:SOURce
and
:CALibration:ADJ:SENSe
commands. For example, a typical value for the 1 mA range
would be:
:CAL:ADJ:SOUR -1.025e-3
:CAL:ADJ:SENS -1.025e-3
Set the current source to 0 A output using the
:SOURce:CURRent
0.0
command.
Record the DMM reading in the "Current calibration adjustment data" table in
(on page 4-4).
Send the source and sense adjustment commands using the DMM reading for the parameter.
For example:
:CAL:ADJ:SOUR 1e-6
:CAL:ADJ:SENS 1e-6
Set the source to the positive full-range value using the
:SOURce:CURRent
command. For
example, for the 1 mA range:
:SOUR:CURR 1e-3
Record the DMM reading in the "Current calibration adjustment data" table in
(on page 4-4).
Send the source and sense commands using the DMM reading as the parameter. For
example:
:CAL:ADJ:SOUR 1.03e-3
:CAL:ADJ:SENS 1.03e-3
Send the
:SOURce:CURRent
command to set the source current to 0 A:
:SOURce:CURRent 0.0
Record the DMM reading in the "Current calibration adjustment data" table in
(on page 4-4).
Send the
:CALibration:ADJ:SOURce
command using the DMM reading as the command
parameter. For example:
:CAL:ADJ:SOUR -1.02e-3