Section 4: Adjustment
Model 2461 Calibration and Adjustment Manual
4-6
2461-905-01 Rev. A / February 2016
Source range
Source current
DMM current reading*
4 A
+4.0000 A
____________________________ A
+000.00 A
____________________________ A
−
4.0000 A
____________________________ A
−
000.00 A
____________________________ A
5 A
+5.0000 A
____________________________ A
+000.00 A
____________________________ A
−
5.0000 A
____________________________ A
−
000.00 A
____________________________ A
7 A
+7.0000 A
____________________________ A
+000.00 A
____________________________ A
−
7.0000 A
____________________________ A
−
000.00 A
____________________________ A
10 A
n/a
There are no adjustment steps for the 10 A
range.
* DMM reading used in corresponding adjustment step. See the adjustment procedure in