Section 4: Adjustment
Model 2461 Calibration and Adjustment Manual
4-10
2461-905-01 Rev. A / February 2016
Voltage range calibration adjustment commands
Step
Command
Description
1
:SOUR:VOLT:RANG <Range>
Select the source range.
2
:SOUR:VOLT -<Range>
Establish negative polarity.
3
DMM reading 1.**
Read the actual output value on the DMM.
4
:CAL:ADJ:SOUR <DMM_Reading1>
Adjust the source function negative full scale.
5
:SYST:ERR?
Check Model 2461 for errors.
6
:CAL:ADJ:SENS <DMM_Reading1>
Adjust the sense function negative full scale.
7
:SYST:ERR?
Check Model 2461 for errors.
8
:SOUR:VOLT 0.0
Set the output to 0 V.
9
DMM reading 2.**
Read the actual output value on the DMM.
10
:CAL:ADJ:SOUR <DMM_Reading2>
Adjust the source function negative zero.
11
:SYST:ERR?
Check Model 2461 for errors.
12
:CAL:ADJ:SENS <DMM_Reading2>
Adjust the sense function negative zero.
13
:SYST:ERR?
Check Model 2461 for errors.
14
:SOUR:VOLT +<Range>
Establish positive polarity.
15
DMM reading 3.**
Read the actual output value on the DMM.
16
CAL:ADJ:SOUR <DMM_Reading3>
Adjust the sense function positive full scale.
17
:SYST:ERR?
Check Model 2461 for errors.
18
CAL:ADJ:SENS <DMM_Reading3>
Adjust the source function positive full scale.
19
:SYST:ERR?
Check Model 2461 for errors.
20
:SOUR:VOLT 0.0
Set the output to 0 V.
21
DMM reading 4.**
Read the actual output value on the DMM.
22
:CAL:ADJ:SOUR <DMM_Reading4>
Adjust the source positive zero.
23
:SYST:ERR?
Check Model 2461 for errors.
* Perform the complete procedure for each range, where
<Range>
= 0.2, 2, 7, 10, 20, and 100.
** The
<DMM_Reading
n
>
parameter is the DMM reading from the previous step.
Each range requires:
Four
CAL:ADJ:SOURce
commands (full scale and zero for each polarity)
Three
CAL:ADJ:SENSe
commands (full scale for each polarity, and zero)
Step 3. Current adjustment
Keithley Instruments recommends two different connection configurations for current adjustment:
For the 1 µA to 1 A ranges, use the connection diagrams in
Connections for the 1 µA to 1 A
(on page 2-5).
For the 4 A to 10 A ranges, use the connection diagrams in
Connections for the 4 A to 10 A
(on page 2-5).
The following topics describe how to do current adjustment for each of these configurations.