Performance tests
10 MHz reference frequency accuracy
Required equipment
Prerequisites
Frequency counter
SMA female-to-BNC male adapter
50
Ω
SMA cable
1.
Connect the 10 MHz Reference Output on the back of the instrument to the
A input of the frequency counter, using the 50
Ω
SMA cable and SMA-BNC
adapter.
Figure 2-12: Equipment connection for verifying the 10 MHz reference frequency
accuracy
2.
On the frequency counter, press the Meas and the Freq buttons.
3.
Verify that the frequency counter reading falls within the range of
9.99999 MHz to 10.00001 MHz (±1 ppm).
4.
Disconnect the test setup.