Speci
fi
cations
Electrical speci
fi
cations
Table 1-1: Run mode
Characteristics
Description
Continuous mode
An arbitrary waveform is output continuously.
Triggered mode
An arbitrary waveform is output only once when a trigger signal is applied. After the waveform
is output, the instrument waits for the next trigger signal.
Triggered continuous mode
An arbitrary waveform is output continuously after a trigger signal is applied.
Table 1-2: Arbitrary waveform
Characteristics
Description
Waveform memory
AWG70001A
Standard
2,000,000,000 points
With Option 01
8,000,000,000 points when
≤
25 GS/s
16,000,000,000 points when >25 GS/s
AWG70002A
Standard
2,000,000,000 points
With Option 01
8,000,000,000 points
Minimum waveform size
AWG70001A
1 point in continuous run mode
4800 points in triggered run modes
AWG70002A
1 point in continuous run mode
2400 points in triggered run modes
Waveform granularity
AWG70001A
1 point in continuous run mode
2 points in triggered run modes
AWG70002A
1 point in continuous run mode
1 point in triggered run modes
Waveform interleaving
Interleaving is performed without zeroing
AWG70001A
Non-interleaved when
≤
25 GS/s, interleaved when >25 GS/s
AWG70002A
Non-interleaved at all sample rates
DAC resolution
8-bit, 9-bit, or 10-bit
1–2
AWG70000A Series and AWGSYNC01 Technical Reference
Summary of Contents for AWG70001A
Page 2: ......
Page 12: ...Service safety summary viii AWG70000A Series and AWGSYNC01 Technical Reference...
Page 14: ...Preface x AWG70000A Series and AWGSYNC01 Technical Reference...
Page 15: ...Specifications...
Page 16: ......
Page 42: ...AWGSYNC01 1 26 AWG70000A Series and AWGSYNC01 Technical Reference...
Page 43: ...Performance verification...
Page 44: ......