Performance tests
Figure 2-16: Equipment connection for checking the analog amplitude accuracy
5.
Press the
Home
button on the instrument, or click the Home tab on the display.
6.
Click the
Reset to default setup
button in the toolbar.
7.
On the instrument, load the
PV_DC_Plus.wfmx
waveform as an output
waveform:
a.
Select
Open Waveform
.
b.
In the dialog box, navigate to the
C:\Program
Files\Tektronix\AWG70000\Samples\PV
folder, and then
select the
PV_DC_Plus.wfmx
fi
le. The
Waveform List
window appears.
c.
In the window, drag and drop the
PV_DC_Plus.wfmx
waveform on the
work space.
8.
Set the amplitude of the instrument as shown in the
fi
rst row of the table:
Table 2-6: AC analog amplitude accuracy
Model
Amplitude settings
Accuracy limits
250 mV
p-p
244 mV to 256 mV
375 mV
p-p
367 mV to 383 mV
AWG70000A series
500 mV
p-p
489 mV to 511 mV
9.
Press the Ch 1
Enable
button on the instrument to enable the channel 1 output.
10.
Press the
Play
button on the instrument to output the signal.
AWG70000A Series and AWGSYNC01 Technical Reference
2–29