Brief procedures
Checking the AC output
(AWG70001A with option
AC)
Required equipment
Prerequisites
Signal analyzer
One Planar Crown RF Input Connector –
7005A-1 SMA Female
One 50
Ω
SMA cable
Two 50
Ω
SMA terminations
None
1.
Press the
All Outputs
button on the instrument to turn off all the outputs.
2.
Click the
Reset to default setup
button in the toolbar.
3.
Use a 50
Ω
SMA cable to connect the AC connector on the instrument to
the RF input of the signal analyzer.
Figure 2-6: Equipment connections for checking the AC output
AWG70000A Series and AWGSYNC01 Technical Reference