Performance tests
11.
Press the
All Outputs On/Off
button on the instrument to output the
waveform.
12.
Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1
.
13.
Use the following formula to compensate the voltage for the 50
Ω
BNC
termination:
V_high = [( 50) / (2 Term_R)] Measured_voltage_1
Where Term_R is the resistance of the 50
Ω
BNC termination measured in
step 3 in the
Measure the termination resistance
procedure. (See page 2-28,
Measure the termination resistance
14.
In the
Waveform List
window, select the
PV_DC_Minus.wfmx
waveform
on the
User De
fi
ned
tab.
15.
Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_2
.
16.
Use the following formula to compensate the voltage for the 50
Ω
BNC
termination:
V_low = [( 50) / (2 Term_R)] Measured_voltage_2
Where Term_R is the resistance of the 50
Ω
BNC termination measured in
step 3 in the
Measure the termination resistance
procedure. (See page 2-28,
Measure the termination resistance
17.
Verify that the voltage difference |(V_high-V_low)| falls within the limits
given in the table. (See Table 2-6 on page 2-29.)
18.
Repeat steps 8 through 17 for each amplitude setting in the table. (See
Table 2-6 on page 2-29.)
19.
Press the
All Outputs On/Off
button to turn off all the outputs.
20.
Disconnect the test setup.
2–30
AWG70000A Series and AWGSYNC01 Technical Reference
Summary of Contents for AWG70001A
Page 2: ......
Page 12: ...Service safety summary viii AWG70000A Series and AWGSYNC01 Technical Reference...
Page 14: ...Preface x AWG70000A Series and AWGSYNC01 Technical Reference...
Page 15: ...Specifications...
Page 16: ......
Page 42: ...AWGSYNC01 1 26 AWG70000A Series and AWGSYNC01 Technical Reference...
Page 43: ...Performance verification...
Page 44: ......