![Tektronix 502A Instruction Manual Download Page 9](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909009.webp)
Operating
Instructions
—
Type
502A
Fig.
2-4.
Inputs to the vertical amplifiers. Waveforms applied to
input
connector A are
displayed
in
the upright
position,
while wave
forms
applied to input connector B are inverted.
Differential
inputs
must be
connected to both the A and B
input connectors
of one
amplifier. In the A and
B positions
of
the input selector switch, the corresponding waveforms
are
displayed
on the crt. In the A-B (DIFF) position, the
B
input
is algebraically added to the A input and the
difference is
displayed. This feature
permits you to elim
inate common
mode
signals
within specified limits.
Input
Coupling
Input
signals to the vertical amplifiers can be either
ac
or dc
coupled by
placing the Input Selector switch in
the
corresponding positions. DC coupling applies both the
ac
and
dc components of the input signal
to the vertical
amplifier
circuits. This permits
you to measure the dc
voltage
level as well as the amplitude
of the ac component.
It
is
sometimes
neither necessary nor desirable to display
the d
c
component.
In such cases ac coupling should
be used.
With
ac coupling,
a capacitor is placed in series with the
input
connector to block the dc component while at the same
time
allowing
the ac component to be displayed.
Fig.
2-5.
Rejection of a common mode signal by the differential
amplifier.
The waveform
applied
to
input connector B is algebra
ically
added to the waveform applied to input
connector A and
the resultant
waveform is displayed on the screen of
the
crt.
Input
Connections
Here
are
some precautions you should observe in connect
ing
your
oscilloscope to
the signal source:
1.
Avoid
errors in readings due to stray electric or magne
tic
coupling between circuits, particularly in the leads
connected
to the input
connector. In general, unshielded
leads
of appreciable length are
unsuited for this use. When
shielded
leads are
used, the shields should be grounded to
the
input connector
shield and
to the chassis of the equipment
being
tested. Coaxial cables are recommended for many
purposes.
Special
care must be taken in the high sensitivity
ranges
of
the oscilloscope due to the low signal level and
high
amplifier
gain.
2.
As nearly as
possible, simulate actual operating condi
tions
in the equipment being tested. The equipment should
have
a load on it which is
approximately equal to the load
encountered
in
normal operation.
3.
Consider the effect
of loading upon the signal
source due
to the input circuit of
the oscilloscope. The circuit at the
input connectors
can
be represented by a resistance of 1
megohm
shunted
by a capacitance of
47 pf. With a few
feet
of
shielded cable, the capacitance may well be 100 pf
or more. In
many
cases, the effects of these resistive and
capacitive
loads are not negligible,
and to minimize them,
you
might
want to use a probe in the manner described in
the next
section.
Use of Probes
An
attenuator probe lessens both capacitive and resistive
loading, at the same time
reducing sensitivity. When making
amplitude
measurements
with an attenuator probe, be
sure
to
multiply
the observed amplitude by
the attenuation of the
probe.
A
Type P6006 Probe is furnished as an
accessory to the
Type
502A Oscilloscope. Connected to the
Input connector
of
the Type 502A,
the
probe presents
an input characteristic
of
10 megohms shunted
by approximately 9 picofarads
and
has
an attenuation ratio
of 10:1. The maximum voltage
which
may be applied to the probe is
600 volts. Exceeding
this rating, either in
peak ac
volts or dc
volts, may result
in
damage
to
the components inside the probe body.
If
the
waveform
being displayed contains fast changing
portions,
it is
generally necessary to clip the ground
lead
of
the probe to the chassis of the equipment being tested.
Select
a ground point which is near the probe connection.
Probe
Adjustment
An
adjustable
capacitor, which
is built into
the probe
body, is used
to compensate for slight variations in input
capacitance
from one instrument to another. To insure the
accuracy
of
pulse and transient measurements, this adjust
ment
should
be checked frequently.
To
make
this adjustment, set the AMPLITUDE CALIBRATOR
and
SENSITIVITY Controls
on the Type
502A to display a
signal of suitable
amplitude.
Touch the probe tip to the
CAL OUT Connector and adjust the TIME/CM
Control to
display
several cycles of the waveform.
Loosen
the
locking sleeve
several turns.
(See figure 2-6.)
Now
while holding onto the knurled part near the cable, turn
2-3
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...