![Tektronix 502A Instruction Manual Download Page 15](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909015.webp)
Operating
Instructions
—
Type
502A
(Do
not
adjust the vertical positioning control after this
reference
point has been established.)
3. Remove the
probe tip from ground and connect the probe
to
the
signal source. Adjust the TRIGGERING LEVEL Control
for
a
stable display.
4. Measure the vertical distance in centimeters from
the
point to
be
measured to the ground
reference line by
use of
the graticule.
5. Multiply the setting of
the SENSITIVITY Control
by the
distance
measured to obtain the indicated voltage.
6.
Multiply
the
indicated
voltage by the attenuation factor
of
the probe
you are using to obtain the true voltage with
respect
to ground.
As
an example of this method, assume that you are us
ing
a 10X probe and a
sensitivity of .2 volts per centimeter
and
that
after setting the reference line at the bottom of the
graticule,
you measure
a
distance of 8 centimeters to the
point
you
wish to measure. In this case then, 8 centimeters
times
.2 volts per centimeter gives you an indicated
voltage
of 1.6
volts.
Since the voltage
point is above the ground
reference
line, the polarity is
indicated to be positive. The
indicated
voltage
times the
probe attenuation foctor
of 10
then
gives
you the true voltage of positive 16 volts.
You
should remember
in determining the polarity of
voltages
measured in this fashion that inputs
applied to
the
B input connectors
are inverted on the face of the crt.
Consequently,
the apparent polarity of these inputs is op
posite
the true polarity.
To prevent possible confusion as
to
polarity it
is usually best to use the A input connectors
for
voltage
measurements with respect
to ground.
Time Measurement
Accurate elapsed time
or time interval measurements
can
be made by utilizing the calibrated timebase feature
of
the Type 502A Oscilloscope.
The sweeps are calibrated
so
that the beams are
deflected across the screen at known
rates. Since
the beam travels completely across the screen
in
a known period of time, the time required for the beam to
travel any portion
of the distance can be determined. By
measuring
the
horizontal
distance between points on the
displayed waveform,
and
by knowing the sweep rates, you
can
determine the time interval between the
two points.
For
accurate measurements be sure the same reference point
is
used
on both pulses and the VARIABLE TIME/CM Control
is
in
the CALIBRATED position.
The method for measuring
a
time interval is as
follows:
1.
Using
the
graticule, measure the horizontal distance
in
centimeters between two points whose time interval you
wish
to
find.
2.
Multiply
the distance measured by the setting of the
TIME/CM
control to
obtain the apparent time interval.
3. Divide the apparent time interval by the magnification
indicated
by the setting of the HORIZONTAL DISPLAY switch
to
obtain
the correct time interval.
Frequency Measurement
The
frequency of a periodically
recurrent waveform can be
determined
if
the time interval (period) of one complete
Fig. 2-12. Measuring
the time interval (period) of one complete
cycle
of a periodically recurrent waveform. The horizontal distance
in
centimeters of one complete cycle
(in this case 3 centimeters) is
multiplied
by the
setting
of the TIME/CM control and is
divided
by
the sweep
magnification to obtain the time interval. The fre
quency of the waveform
is the reciprocal of its time interval.
cycle
of the waveform is known. This time interval can be
measured
by means of
the
procedure described in the preced
ing
paragraph.
The frequency of a waveform is the recipro
cal
of
its
time interval
Using
DC BAL and FINDER Controls
It is best
to
disconnect the input signal, unless the
approximate
amplitude of it is known. If the amplitude of
the
input
signal is known then set
the SENSITIVITY Control
for a
crt display of
two
or three centimeters and follow
the
instructions for no input signal.
With
no
input signal, turn the
TRIGGERING LEVEL to
RECURRENT,
the single sweep MODE
switch to
NORMAL
and
the INTENSITY
Control to a reasonable level.
The above
control
settings are
to insure that a trace can be seen, when
it is found.
Set
the vertical POSITION
Control of the proper beam to
its
midrange position. With the SENSITIVITY Control in the
.2
VOLTS
PER CM position. If no input signal is connected to
the
Type
502A, change
the SENSITIVITY Control to
the .1 m
VOLTS
PER CM position and depress the FINDER Button. The
trace should
now
be seen on the crt.
Now
turn DC BAL
Control in either
direction
until a
"fast area” is seen.
A 'fast
area”
is an
area of the DC BAL Control where the trace will
move
a
large distance for
a very small movement of the DC
BAL
Control.
When
the
"fast
area" of the
DC BAL Control is found,
the
trace should
be positioned in the center of
this area
with
the
DC BAL
Control.
The FINDER Button may now be
released and the trace
positioned
to the proper place on
the
graticule.
2-9
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...