Calibration
Procedure
—
Type
502A
13. Coaxial "Straight" Connector. (1)
14. Jumpers,
6
inch with banana
plugs. (3)
15.
A constant amplitude
sine-wave generator having
output
frequencies
of 1 kilocycle,
50
kilocycles, 100 kilocycles, 200
kilocycles,
400 kilocycles and
1 megacycle. The
generator
should also
have
a variable output amplitude from 500
μvolts
to 2 volts.
If a constant
amplitude generator is not available to
cover
the above ranges, any generator which has the
proper
frequencies
and output amplitudes may be used.
When
using a
generator which does
not have a constant
amplitude
feature
the test oscilloscope should
be used
to
monitor the output of the generator. By use of the
oscilloscope
the
output signal from the generator may be
kept
at
a constant amplitude.
By
the use of a DB
box or an attenuator system, the
signal
going
from the sine-wave generator to the Type 502A
may
be reduced to the proper
amplitude for the
high sensi
tivity
ranges,
while at the
same time the signal going from
the
sine-wave generator to the test oscilloscope may be
kept
at
a high
amplitude to enable monitoring for a constant am
plitude
of signal
out of the sine-wave generator.
Preliminary
NOMINAL
RESISTANCES OF
POWER
SUPPLIES
SUPPLY
APPROX.
RESISTANCE TO
GROUND
-150V
3.5
k
100
V
4
k
350V
10k
Connect
the
power cord
and the ac voltmeter to the out
put
of the autotransformer. Rotate the POWER AND SCALE
ILLUM
control clockwise to the ON position and adjust the
autotransformer
for an output of 117 volts (or 234 volts).
Allow
the
instrument to warm up for several minutes
before
proceeding
with the calibration adjustments. During cali
bration,
periodically check the input voltage to the instru
ment
and adjust the autotransformer as necessary to
main
tain
the voltage
at 117 or 234 volts except when the power
supply
regulation is being checked.
CAUTION
Do
not
reset
the —150 control unless the power
supply
voltages are actually out of tolerance or
you are
planning
to perform a complete calibra
tion
of the instrument.
Preset
the front-panel controls
of
the
Type
502A
as
follows:
POWER
AND SCALE ILLUM
PWR
OFF
INTENSITY
full
counterclockwise
INTENSITY
BALANCE
midrange
FOCUS (both beams)
midrange
TRIGGERING
LEVEL
midrange
TRIGGER
SELECTOR
+
LOWER AC
MODE
Switch
NORMAL
TIME/CM
1
mSEC
VARIABLE
CALIBRATED
HORIZONTAL
DISPLAY
2
EXT VOLTS/CM
HORIZONTAL
POSITION
midrange
AMPLITUDE
CALIBRATOR
OFF
HORIZ
DEF PLATE
TIME
BASE AMP
SELECTOR
The
following applies to both vertical amplifiers:
Input Selector
A-DC
SENSITIVITY
.2
VOLTS PER CM
VARIABLE
CALIBRATED
DC
BAL.
control
will be
set later.
VERTICAL
POSITION
midrange
Before
applying power to the instrument
the resistances
of
the power supplies should be checked. The typical resis
tances
of
the supplies may be found in the chart below.
1.
Low-Volfage Power
Supply
Connect
the dc voltmeter to the
output of the —150 volt
power supply. Power supply output
test
points are shown in
Fig.
5-2. Set the
—150 Control for a reading of —150 volts.
Check that
the outputs of the —6.2, +100 and
+350 volt
supplies are within the
spe or — 3% tolerances.
If either the
+100 or +350 volt supplies are out of tolerance,
reset
the
—150 control slightly until all power
supply
voltages
are
within tolerances.
Vary
the output of the autotransformer between 105 and
125
volts
(or between 210 and 250 volts) to check the regula
tion
of
the low voltage power supplies. The power supplies
must
regulate within
+ or — 3% of their correct voltages.
Fig.
5-2. Power Supply voltage test points.
5-2
®2
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...