![Tektronix 502A Instruction Manual Download Page 48](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909048.webp)
Calibration
Procedure—
Type
502A
15.
Gas and Hum
Check
With jumpers,
ground
both the inputs and set the SENSI
TIVITY
switch at .2 mVOLTS
PER
CM. Turn TRIGGERING
LEVEL
to RECURRENT
and using the POSITION and DC BAL
Controls,
position the trace onto the crt. The use of the DC
BAL
Control
may be
found in Section Two.
Now
turn
the Input Selector
switch from A-DC to A-AC
and
B-DC to B-AC. The trace should not have shifted more
than
2 cm during the switching from DC to AC.
To
check for hum, observe the amount of hum on the free
running
trace. It should not be more than about 3 mm.
16.
.2V/CM DIFF BAL
Set the
AMPLITUDE
CALIBRATOR
switch at .5V and
apply
the calibrator output to
both A and B input connectors
of
the
vertical amplifier. Place the SENSITIVITY Control in
the
.2
VOLTS
PER CM position and the Input Selector switch
in
the A-B (DIFF) DC
position. Set
the TIME/CM switch at
1
mSEC. Position
the trace so that it lies behind the appro
priate
horizontal centerline.
Adjust the .2 V/CM DIFF BAL
Control
for minimum vertical deflection.
A-DC
position
and the SENSITIVITY switch in the .2 VOLTS
PER
CM position. Set the .2V/CM GAIN for 2.5 centimeters
of
vertical
deflection.
17. .2MV/CM DIFF
BAL
Place
the SENSITIVITY
switch in the .2 mVOLTS PER CM
position
and position the trace so that it lies behind the
appropriate
horizontal
centerline. Adjust the .2MV/CM DIFF
BAL Control
for
minimum
vertical
deflection. Repeat steps
16 and
17
until both adjustments are correct.
18.
.2V/CM GAIN
Apply
a .5 volt signal from the calibrator to input A of
the vertical amplifier.
Set the
Input Selector switch in the
Fig.
5-8. Setting
the .2 V/CM GAIN
control. In the calibration
procedure,
the
.2 V/CM GAIN control is adjusted for 2.5 centi
meters
of
vertical deflection.
NOTE
To
trigger
the display properly set the
TRIGGER
SELECTOR
to the +AC position for the beam
which
is being
calibrated.
Fig.
5-7.
Adjustment of the .2 MV/CM DIFF. BAL. control. If the .2MV/CM DIFF. BAL. control is slightly misadjusted, a waveform similar
to waveform A
will
be obtained. Waveform
B shows a typical display obtained when the control is adjusted properly.
5-6
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...